Setting noise filtering

Noise filtering is used when waveform breaks appear in the profile. Noise components,
the cause of waveform breaks, can be filtered out.
MEAS/NOISE
[0]
OK
FUN mode-[MEAS]-[IMAGE]-[PROFILE]-[NOISE]
Setting value
0 to 7 (pixel)
Received Light Gain
This is used when insufficient light amount prevents profiles from being displayed
correctly. The received light gain can be changed in stages up to eight times.
Setting of the received light gain is enabled only in the EXP mode.
FUN mode-[MEAS]-[IMAGE]-[GAIN]
Setting value
LV1, LV2 (1.5 times), LV3
(2 times), LV4 (3 times),
LV5 (4 times), LV6 (6
times), LV7 (8 times)
82
Measurement Settings
MEAS/NOISE
[5]
OK
Description
Light received signals of width smaller than the specified size is
filtered as noise. An optimum value is set according to the Sensor
installation status (regular reflection or diffuse reflection) as the
default value.
Description
Any received light gain up to eight times can be set up. (defaults: in
standard mode or high-speed mode, LV1, in high-resolution mode,
LV2)