Setting the firmware information
Updating the firmware
198
Updating the Firmware
6
Click the [Controller information]
button if necessary.
The model and version of the
currently connected Controller are
displayed.
7
Click the [Loading file] button to
select the file to be written.
The model and version of the
Controller that is held in the file are
displayed.
8
Click the [Start upgrade] button.
A message indicating the start of the
update is displayed.
Important
If the message "the model is not the
same" is displayed when you click the
[Start upgrade] button, this means that
the model of the connected Controller
and the model information in the
specified file do not match. In this
case, do not update the firmware. The
Controller will break down and can no
longer start up normally.