Mounting the ZG-WDS8T
Fix by mounting screws making sure that the distance between the Sensor and
measurement target is matched.
Mounting method
Mounting position
Mount the Sensor according to the following distances and angle.
Emission axis
50 mm
Measurement target
Mounting for diffuse reflection
measurement
Note
The default mounting state of the Sensor is for diffuse reflection measurement. To set the
Sensor for regular reflection measurement, change the Sensor mounting setting.
Setting the Sensor Installation Status p.100
26
Mounting and Connecting Devices
Reception axis
30°
Fasten the Sensor onto the mounting base
with M4 screws.
Tightening torque: 1.2 N•m
Important
For details on the positions of screw holes,
check the external dimensions in "Chapter 5
APPENDICES."
External dimensions p.164
15°
Emission axis
Reception axis
44 mm
Measurement target
Mounting for regular reflection
measurement