~CWARCS
A419-00-880 Issue K
Figure
6
EPX
N series dry
pump
Table 5-
EPX
N (refer to Figure 6)
Number
Description
1
Pipe fitting · Nitrogen supply to the seal purge.
2
Label ·
Warning;
If the pump has overheated or the cooling water has failed, connecting
cooling water to a hot pump will cause severe damage to the pump.
Always allow the pump to cool for
at
least 20 minutes before connecting the cooling water.
3
Label ·
Warning;
do not remove cover until 4
minutes
after disconnecting power from the
pump.
4
Gas supply fitting· Connect the gas supply to this connection. This compression fitting accepts
1/
4 inch (6.35
mm)
diameter tube. Ensure t hat the tube used is suitable for
the
application.
5
Label · Gas Module identifier.
6
Pipe fitting · Nitrogen supply to the gas ballast.
©