Table 18
-
EPX
EUC alarms and warnings
......
Message on PDT
Description
Action
ALARM 101
EUC controller fault.
Cycle power off and on. If fault still
present,
Watchdog Reset
contact Edwards representative
ALARM
101
The emergency trip circuit has
triggered,
Reset the emergency stop
circuit.
STOP ACTIVATED
and the inverter is disabled.
Allow the pump to stop completely. Restart
the pump.
WARNING 101
A power loss of more than half a second
Restore power. Restart the pump.
POWER INTERRUPT
has occurred. This warning is shown as
the power is removed.
WARNING 113
General
Input/
Output fault on EUC.
Cycle the power off and on. If fau lt still
PCA FAULT
present,
contact Edwards representative.
WARNING 312
Inverter has indicated over-current.
Monitor the
pump.
If the problem
persists,
CURRENT HIGH
stop the pump, remove power and contact
Edwards representative.
ALARM
501
Inverter has received a Pump Thermal
Remove power to the pump. Check that
PUMP TEMP HIGH
Overload input.
cooling water flow rate is at least 2
1/
min
and other services are correctly supplied.
Allow the pump to cool down completely.
Restart the pump. If the problem
persists,
contact Edwards representative.
WARNING 3501
Nitrogen flow
low;
need to check gas
Check the nitrogen flow rate to the pump.
N2 FLOW LOW
supply.
Ensure that supply exceeds 17 slm. If problem
persists,
contact Edwards representative.
ALARM
7001
Inverter running
signal
has not been
Remove power. Check that the power supply
INVERTER STOPPED
received for more than 20 seconds whilst is at full
strength.
Apply power and restart
t
he pump should be
running.
the pump. If the problem
persists,
contact
Usually caused by a dip in the 3-phase
Edwards representative.
supply.
ALARM 17601
Inverter has indicated a major
fault/
trip Remove power to the pump. Check that
INVERTER FAULT
condition across the serial link.
cooling water flow rate is at least 2
I/
min
and other services are correctly supplied.
Allow the pump to cool down completely.
Restart the pump. If the problem
persists,
contact Edwards representative.
ALARM 17601
Inverter has not indicated
'Running'
Remove power to the pump. Check that
NO RUN SIGNAL
across the serial link when commanded
cooling water flow rate is at least 2
1/
min
'On'.
and other services are correctly supplied.
This will occur after a
thermal
trip if the Check exhaust line is not blocked and pump
pump has not been allowed to cool for
inlet is clear.
long enough.
Allow the pump to cool down completely.
Restart the pump. If the problem
persists,
contact Edwards representative.
WARNING 17601
Inverter has indicated a minor fault/trip Remove power to the pump. Check that
INVERTER FAULT
condition across the serial link.
cooling water flow rate is at least 2
I/
min
and other services are correctly supplied.
Allow the pump to cool down completely.
Restart the pump. If the problem
persists,
contact Edwards representative.
WARNING 17613
Lost communication with inverter.
No action is required unless the pump is
INVERTER FAULT
stopped,
in which case, remove the power to
the pump (for
example,
by operating the EMS
circuit).
©