riicWARCS
A419-00-880 Issue K
Table 3
EPX
L
pump labelling information (Refer to Figure 4)
Number
Label Name
Description
1
Pump details
This provides specific information about the pump including: pump
type;
code number; serial
number;
pump
weight;
year of
manufacture;
and name and address of the manufacturer. It also
includes electrical supply requirements, and compliance marks.
2
Power
The power LED. See
Table 1
for details.
3
Run
The run LED. See
Table 1
for details.
4
Warning
The warning LED. See
Table 1
for details.
5
Alarm
The alarm LED. See
Table 1
for details.
6
Warning · risk of electric
Do not remove cover until4 minutes after disconnection of the
shock
power.
7
Warning · risk of electric
Do not remove cover until4 minutes after disconnection of the
shock
power.
8
Status LED label
Status LED label.
9 and 10
Lifting point
The pump must be lifted using the eyebolts on the pump and
suitable lifting equipment.
11
Warning · risk of high
Surface temperatures can exceed 65
C during pump operation.
temperature
12
Direction of rotation
Arrow
shows the correct direction of rotation of the rotor.
13
Warning · danger of injury
Turn off and lock-out all power prior to servicing.
from rotating parts
14
Water in
The cooling water supply pipeline connects to the pump at this
location.
15
Protective earth stud
This equipment must have a second protective earth. See
Section
3.10.
16
Pump interface
Tool Interface Module (TIM) fits in this position. The pump may be
controlled through this. See
Section 3.11
.
17
Warning · risk of electric
Do not remove cover until 4 minutes after disconnection of the
shock
power.
18
Warning · risk of electric
Do not remove cover until 4 minutes after disconnection of the
shock
power.
19
Warning · thermal shock
If the pump has overheated or the cooling water has failed
,
connecting cooling water to a hot pump will cause severe damage to
the pump.
Always allow the pump to cool for at least 20 minutes before
connecting cooling water.
20
Water out
The cooling water return pipeline connects to the pump at this
location.
21
Product label
Product name.
©
Edwards Limited
2017.
All rights reserved.
Page 9