A371–22–880 Issue R Original Instruction Manual E2M0.7 and E2M1.5 Rotary Vacuum Pumps Description Item Number E2M0.7 (0.7 m , 0.4 ft 220-240 V, 50/60 Hz A371-31-919 E2M0.7 (0.7 m , 0.4 ft 100-120 V, 50/60 Hz A371-31-902 E2M1.5 (1.5 m , 0.8 ft...
Crawley, West Sussex, RH10 9LW, UK declare under our sole responsibility, as manufacturer and person within the EU authorised to assemble the technical file, that the product(s) E2M0.7 and E2M1.5 Rotary Vacuum Pumps without IEC connector A371-31-919 A371-22-919 A371-13-930 A371-31-902...
Introduction Scope and definitions This manual provides installation, operation and maintenance instructions for the Edwards E2M0.7, and E2M1.5 rotary vacuum pumps. You must use the pump as specified in this manual. Read this manual before you install and operate the pump.
The E2M0.7 and E2M1.5 pumps are two-stage, direct drive, sliding vane pumps. The pump is oil-sealed and designed for reliable, long-term operation. The pump is a free-standing unit. The drive is provided through a coupling by a single-phase motor.
A371–22–880 Issue R Lubrication data Note: An Edwards Safety Data Sheet for Ultragrade 15 is available on request. Table 5 - Lubrication data Recommended oil Ultragrade 15 Maximum oil capacity 0.32 litre To operate the pump when the ambient temperature is outside the range specified in Section 2.1, or to...
You must ensure that the pump is suitable for your application, If you have any doubts as to the suitability of the pump for your application, refer to the Edwards guidelines on vacuum pump and vacuum system safety (see the Associated publication at the end of the contents list at the front of this manual).
0.5 bar gauge (1.5 bar absolute, 1.5 x 10 Pa, 21.75 psi) at full pump throughput. In the following circumstances, Edwards recommend an oil mist filter is fitted to the pump outlet if the pump is used with the gas-ballast control open ...
2.2, make sure that this is not due to your system design before you contact your supplier or Edwards for advice, In particular, the vapour pressure of all materials used in your vacuum system (including pump oil, see below) must be much lower than the specified ultimate vacuum of the pump.
The pump is designed for unattended operation under the normal operating conditions specified in Section However, Edwards recommend checking the pump at a regular interval of not more than 14 days; check the pump more frequently if pumping high volumes of gas or if operating the pump with the gas-ballast control open.
Fluorinated materials in the pump may include oils, greases and seals. The pump may have overheated if it was misused, if it malfunctioned or if it was in a fire. Edwards Material Safety Data Sheets for fluorinated materials used in the pump are available on request: contact your supplier or Edwards.
The majority of these centres employ Service Engineers who have undergone comprehensive Edwards training courses. Order spare parts and accessories from your nearest Edwards company or distributor. When you order, state for each part required: Model and Item Number of your equipment ...