Triggering the Oscilloscope
8.7.7
Timeout Trigger
In Timeout trigger, the oscilloscope triggers when the time interval (∆T) (the time
from when the rising edge (or falling edge) of the input signal passes through the
trigger level to the time from when the neighboring falling edge (or rising edge)
passes through the trigger level) is greater than the preset timeout value, as shown in
Figure 8.13
.
Time Out < T
T
Time Out
Figure 8.13 Timeout Trigger
Trigger Type
Click or tap the Type drop-down button to select "Timeout".
Figure 8.14 Timeout Trigger Menu
After a trigger type is selected, the current trigger setting information (including the
trigger type, source, and level) is displayed in the trigger information label at the top
of the screen, as shown in the figure below. The information will change based on the
trigger settings.