Source
. The selected trigger source is indicated in the trigger information label at the
top of the screen.
Only when you select the channel that has signal inputs as the trigger source, can you
obtain a stable trigger.
Trigger Condition
Set the desired trigger condition in the When item.
Start: triggers when SDA data transitions from high level to low level while SCL
is high level.
Stop: triggers when SDA data transitions from low level to high level while SCL is
high level.
Restart: triggers when another start condition occurs before a stop condition.
MissedAck: triggers when ACK is 1.
Address: the trigger searches for the specified address value. When this event
occurs, the oscilloscope will trigger on the read/write bit. After this trigger
condition is selected:
-
-
-
Data: the trigger searches for the specified data value on the data line (SDA).
When this event occurs, the oscilloscope will trigger on the clock line (SCL)
transition edge of the last bit of data. After this trigger condition is selected, you
can set the following parameters.
-
-
-
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Click or tap the drop-down button of Direction to select "Write", "Read", or
"R/W".
This setting is not available when AddrBits is set to "8 Bits".
Click or tap the drop-down button of AddrBits to select the desired
address bits. The available address bits are "7 Bits", "8 Bits", and "10 Bits".
Click or tap the input field of Address, and then use the pop-up numeric
keypad to set the address of I2C trigger. You can also use the corresponding
multipurpose knob to set the value.
Click or tap the drop-down button of AddrBits to select the desired
address bits. The available address bits are "7 Bits", "8 Bits", and "10 Bits".
Click or tap the input field of Bytes, and then use the pop-up numeric
keypad to set the length of the data. You can also use the corresponding
multipurpose knob to set the value. Its range is from 1 to 5.
Click or tap the input field of Data, and then the "Format" interface is will
displayed. You can select "Bin" or "Hex" data format.
Triggering the Oscilloscope
DHO800 User Guide
99