Voltage Parameters; Figure 10.3 Voltage Parameters - Rigol DHO900 User Manual

Hide thumbs Also See for DHO900:
Table of Contents

Advertisement

Measurements
10.2.1.4

Voltage Parameters

Vmax
Vmin
1. Vmax: indicates the voltage value from the highest point of the waveform to the
GND.
2. Vmin: indicates the voltage value from the lowest point of the waveform to the
GND.
3. Vpp: indicates the voltage value from the highest point to the lowest point of the
waveform.
4. Vtop: indicates the voltage value from the flat top of the waveform to the GND.
5. Vbase: indicates the voltage value from the flat base of the waveform to the GND.
6. Vamp: indicates the voltage value from the top of the waveform to the base of
the waveform.
7. Vupper: indicates the actual voltage value that corresponds to the threshold
maximum value.
8. Vmid: indicates the actual voltage value that corresponds to the threshold middle
value.
9. Vlower: indicates the actual voltage value that corresponds to the threshold
minimum value.
10. Vavg: indicates the arithmetic average value on the whole waveform or in the
gating area. The formula is shown as follows:
DHO800 User Guide
132
Overshoot
Vtop
Vpp
Vamp
Vbase
Preshoot

Figure 10.3 Voltage Parameters

n
x
i
i
=
1
=
Average
n
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Vupper
Vmid
Vlower