Page 2
• RIGOL products are covered by P.R.C. and foreign patents, issued and pending. • RIGOL reserves the right to modify or change parts of or all the specifications and pricing policies at the company's sole decision. • Information in this publication replaces all previously released materials.
Do not expose the battery (if available) to high temperature or fire. Keep it out of the reach of children. Improper change of a battery (lithium battery) may cause an explosion. Use the RIGOL specified battery only. • Handle with Caution.
1,000,000 wfms/s (in UltraAcquire Mode), 50 Mpts memory depth, 12 bits resolution, and low noise. The DHO900 series supports 16 digital channels. One instrument can make an analysis on both the analog and digital signals to meet the embedded design and test scenarios.
This manual gives you a quick overview of the front and rear panel, user interface as well as basic operation methods of DHO900 series. http:// For the latest version of this manual, download it from RIGOL official website ( www.rigol.com Publication Number...
Page 25
Vertical Scale Vertical SCALE Knob Content Conventions in this Manual DHO900 series digital oscilloscope includes the following models. Unless otherwise specified, this manual takes DHO924S as an example to illustrate the operation methods of DHO900 series. Model Max. Analog Bandwidth No.
The consigner or carrier shall be liable for the damage to the instrument resulting from shipment. RIGOL would not be responsible for free maintenance/rework or replacement of the instrument. 2. Inspect the instrument In case of any mechanical damage, missing parts, or failure in passing the electrical and mechanical tests, contact your RIGOL sales representative.
Output DC, 12 V, 4 A, 48 W CAUTION The power adaptor provided in accessories can only be used to power RIGOL instruments. Do not use it for mobile phone and other devices. WARNING To avoid electric shock, ensure that the instrument is correctly grounded.
Setup > Language to select the system language. 4.3.5 To Connect the Probe RIGOL provides passive probes and logic probes (optional) for DHO900 series. For DHO900 Data Sheet. For detailed technical specific probe models, refer to information of the probes, please refer to the corresponding Probe User Guide.
Memory depth refers to the number of points that the oscilloscope can store in a single trigger acquisition. It reflects the storage capability of the acquisition memory. DHO900 series oscilloscope provides up to 50 Mpts memory depth (standard). Trigger Point...
In the trigger debugging process of video signals, the frequency in different part of the signal can be reflected by a different brightness, as RIGOL's digital oscilloscope provides the intensity graded color display function. Experienced users can quickly judge the signal quality and discover abnormalities during the debugging process.
Measurements Measurements DHO900 series oscilloscope provides the quick measurements after "Auto" is selected, auto measurements for 41 waveform parameters, as well as the cursor measurement function. 10.1 Auto Scale When the oscilloscope is correctly connected and has detected a valid input signal, click or tap the function navigation icon >...
Digital Channel Digital Channel DHO900 series oscilloscope provides a standard logic analyzer (LA) function with 16 digital channels. The default channel label is D0-D15. The oscilloscope compares the voltages acquired in each sample with the preset logic threshold. If the voltage of the sample point is above the threshold, it will be stored as logic 1;...
1. File Header There is only one file header in a binary file. The file header contains the following information. Table 23.2 File Header Two-byte characters, RG, indicating that the file is the RIGOL Cookie binary data file format. Version Two-byte, indicating the file version.
PC. The remote control can be realized by using SCPI (Standard Commands for Programmable Instruments) commands. This chapter will illustrate how to use the RIGOL Ultra Sigma software to remotely control the instrument via various interfaces. CAUTION Before connecting the communication cable, please turn off the instrument to avoid causing damage to the communication interfaces.
PC via the USB interface. You can also click USB-TMC to search for the resource. 3. View the device resource The resources found will appear under the "RIGOL Online Resource" directory, and the model number and USB interface information of the instrument will also be displayed.
Page 275
4. View the device resource The resources found will appear under the "RIGOL Online Resource" directory. 5. Control the instrument remotely Right-click the device resource name and select "SCPI Panel Control" to open the remotely command control panel.
After restarting the instrument, insert the USB storage device again to check whether it can work normally. d. If the USB storage device still cannot work normally, please contact RIGOL. 4. The touch-enabled operation does not work. a. Check whether you have locked the touch screen. If yes, unlock the touch screen.
Recommended Accessories 16-channel Logic Analyzer Probe PLA2216 NOTE For all the mainframes, accessories and options, please contact the local office of RIGOL. 27.2 Appendix B: Warranty RIGOL TECHNOLOGIES CO., LTD. (hereinafter referred to as RIGOL) warrants that the product mainframe and product accessories will be free from defects in materials and workmanship within the warranty period.
Appendix RIGOL be liable for any consequential, indirect, ensuing, or special damages for any breach of warranty in any case. 27.3 Appendix C: Factory Settings Press the front-panel or click/tap Default on the Toolbar and a prompt message "Restore default settings?" is displayed. Click or tap OK to restore the instrument to its factory default settings, as shown in the table below.