Memory Depth; Figure 7.3 Memory Depth - Rigol DHO900 User Manual

Hide thumbs Also See for DHO900:
Table of Contents

Advertisement

Waveform Leakage: If the sample rate is too low, the waveform reconstructed
from the sample data does not represent the original signal correctly.
7.4

Memory Depth

Memory depth refers to the number of points that the oscilloscope can store in a
single trigger acquisition. It reflects the storage capability of the acquisition memory.
DHO800 series oscilloscope provides up to 25 Mpts memory depth (standard).
The following formula shows the relations among memory depth, sample rate, and
horizontal time base:
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Trigger Point
Pre-Sample
Memory Depth

Figure 7.3 Memory Depth

T
Delayed Sample
Acquisition System
DHO800 User Guide
55