Trigger Mode
In Sweep menu, select Auto, Normal, or Single as the trigger mode. For details, refer
to descriptions in
Trigger Parameter Setting
Sets the trigger parameters (trigger holdoff and noise rejection) under this trigger
type. For details, refer to descriptions in
Trigger Level
Tap the "Trigger" icon on the small screen, then rotate the knob
right part of the small screen or drag the adjustment icon under the Level menu to
adjust the trigger level. For details, refer to descriptions in
trigger level is displayed in the trigger information label at the top of the screen.
8.7.7
Timeout Trigger
In Timeout trigger, the instrument triggers when the time interval (∆T) (the time from
when the rising edge (or falling edge) of the input signal passes through the trigger
level to the time from when the neighboring falling edge (or rising edge) passes
through the trigger level) is greater than the preset timeout value, as shown in
8.13
.
Trigger Type
Click or tap the trigger information label at the top of the screen, and then click or tap
the drop-down button of Type to select "Timeout" from the drop-down list. Then set
the parameters for the Timeout trigger in
Copyright ©RIGOL TECHNOLOGIES CO., LTD.
All rights reserved.
Trigger Mode
.
Trigger Holdoff
Figure 8.13 Timeout Trigger
To Trigger the Oscilloscope
Noise Rejection
and
Trigger Level
Figure 8.14
.
.
at the upper-
. The current
Figure
DS70000 User Guide
91