Connect the Test Equipment

Connect the 16910/11A Logic Analyzer Pod to the 8133A Pulse
Generator
1 Connect a Transition Time Converter (if required—see page 28) to each of
the four outputs of the 8133A pulse generator except Channel 1 OUTPUT.
2 Connect the SMA/Flying Lead test connector (see "To Assemble the SMA/
Flying Lead Test Connectors" on page 22) with the 50 ohm terminator to
the Transition Time Converter at the 8133A pulse generator Channel 1
OUTPUT. (If Transition Time Converters are not required, connect the
SMA/Flying Lead test connector directly to the pulse generator output.)
3 Connect the two SMA/Flying Lead test connectors without 50 ohm
terminators to the Transition Time Converters at the 8133A pulse
generator Channel 2 OUTPUT and Channel 2 OUTPUT. (If Transition Time
Converters are not required, connect the SMA/Flying Lead test connectors
directly to the pulse generator outputs.)
4 Connect an E5383A Flying Lead Probe Set to Pod 1 of the 16910/11A logic
analyzer.
Chapter 3: Testing Logic Analyzer Performance
Connect the Test Equipment
33