Agilent Technologies 16962A Service Manual page 37

Logic analyzer
Table of Contents

Advertisement

3
Testing Logic Analyzer Performance
N O T E
36
Connect the other E5382A Flying Lead Probe Set to Pod 3
5
of the 16962A logic analyzer. This probe set is used for
the clock input.
Connect the Pod 3 E5382A Flying Lead Probe Set's CLK
6
lead to the pin strip of the SMA/Flying Lead connector at
the pulse generator's Channel 1 OUTPUT.
ground clip
Be sure to use the black ground clip (supplied with the E5382A Flying Lead
Probe Set) and orient the leads so that the black clip is connected to one of
the SMA/Flying Lead connector's ground pins!
Connect the Pod 3 E5382A Flying Lead Probe Set's CLK
7
lead to the SMA/Flying Lead connector at the pulse
generator's Channel 1 OUTPUT. Again, be sure to use the
black ground clip and orient the leads so that the black
clip is connected to ground.
Connect the Pod 1 E5382A Flying Lead Probe Set's bits 2
8
and 10 to the SMA/Flying Lead test connector's pin strip
connector at the pulse generator's Channel 2 OUTPUT.
9
Connect the Pod 1 E5382A Flying Lead Probe Set's bits 6
and 14 to the SMA/Flying Lead test connector's pin strip
connector at the pulse generator's Channel 2 OUTPUT.
16962A Logic Analyzer Service Guide

Advertisement