16962A Logic Analyzer Service Guide
strip, and the center conductors of each SMA
connector.
b
Ensure that there is continuity between each of the
two pins on the right side of the pin strip, and the
SMA connector frames.
c
Ensure that there is NO continuity between the SMA
connector center conductor and the SMA connector
frame (ground).
Finish creating the test connectors:
5
Attach an SMA m- m adapter to one end of each of the
a
four SMA/Flying Lead test connectors.
Attach a 50 ohm terminator to the other end of two of
b
the SMA/Flying Lead test connectors.
Testing Logic Analyzer Performance