Table 8
DSO90604A Oscilloscope Setup
Setup: Channel 1
On
Scale: 50 mV/div
Offset: 500 mV
Skew: (Set later. See
page
38.)
Setup: Horizontal
Scale: 200 ps
Position: 0 s
Reference: Center
External 10 MHz Reference
Clock: not selected
Zoom: not selected
Measure: Markers
Mode: Manual placement
All else: (n/a)
16962A Logic Analyzer Service Guide
Set up the oscilloscope.
3
Set up the oscilloscope according to one of the
a
following tables.
Setup: Channel 2
On
Scale: 50 mV/div
Offset: 500 mV
Skew: 0.0 seconds
Setup: Trigger
Mode: Edge
Source: Channel 1
Level: 500 mV
Edge: Rising Edge
Sweep: Auto
Testing Logic Analyzer Performance
Setup: Channel 3
Off
Setup: Acquisition
Sampling Mode: Real Time,
Normal
Memory Depth: Automatic
Averaging: Enabled
# of Averages: 16
Sampling Rate: Automatic
Sine(x)/x Interpolation:
checked
Setup: Channel 4
Off
Setup: Display
Waveforms
Connect dots
Color Grade: not selected
Infinite Persistence: not
selected
Waveform Brightness: as
preferred
Grid: On, Quantity: 1
Intensity: as preferred