Testing Logic Analyzer Performance

Test Pod 2

54
Test the complement of the bits
Now test the logic analyzer using inverted levels (in other
words, complement data).
1
On the pulse generator, in the PULSE setup for CHANNEL
2, select inverted levels:
On the 81134A pulse generator, select Levels: Inverted.
On the 8133A pulse generator, select COMP (LED on).
Verify the DC offset and adjust it if necessary. See
2
page
37.
3
Adjust the sampling positions (run Eye Finder). See
page
46.
Determine pass or fail (1 of 2 tests). See
4
Switch to the Listing window by selecting the Listing tab
5
at the bottom of the main logic analyzer window.
6
Select the Run Repetitive icon
7
Determine pass or fail (2 of 2 tests). See
1
Disconnect the E5382A Flying Lead Probe Set from Pod 1
and connect it to Pod 2 of the logic analyzer. (The clock
input on Pod 3 remains the clock input when testing
Pod 2.)
On the pulse generator, in the PULSE setup for CHANNEL
2
2, return the outputs to normal (non- inverted or
non- complement) levels.
3
Verify the DC offset and adjust it if necessary. See
page
37.
In the Overview window, select SetupBus/Signal... from
4
the module's drop- down menu.
5
Scroll to the right and unassign all Pod 1 bits.
6
Set the Pod 2 threshold to 500 mV (just as you did for
Pod 1 on
page
43).
Assign bits 2, 6, 10, and 14 of Pod 2.
7
page
50.
.
page
53.
16962A Logic Analyzer Service Guide