N O T E
16962A Logic Analyzer Service Guide
Determine PASS/FAIL (2 of 2 tests)
Pass/Fail Point: The Listing window is set up to search for
the appropriate number of A's and 5's in the acquisition. If
the logic analyzer does not detect the correct number of A's
and 5's, an error window will appear.
Select the Run Repetitive icon
1
run for about one minute. The analyzer will acquire data
and the Listing Window will continuously update.
If the "can't find occurrence" window appears, then the
logic analyzer fails the test.
Check your test setup. If the failure is not the result of a
problem with the test setup, record the failure in the
"Maximum State Data Rate" section of the Performance Test
Record.
Be sure that the black ground clip is making good contact with the ground
pin on the test connector.
After about one minute, select the Stop button
2
the acquisition.
If the "can't find occurrence" window does not appear,
then the analyzer passes the test. Record "Pass" in the
"Maximum State Data Rate" section of the Performance Test
Record.
Testing Logic Analyzer Performance
. Let the logic analyzer
3
to stop
53