Agilent Technologies 16962A Service Manual page 35

Logic analyzer
Table of Contents

Advertisement

3
Testing Logic Analyzer Performance
Table 9
54855A Oscilloscope Setup
Setup: Channel 1
On
Scale: 50 mV/div
Offset: 500 mV
Skew: (Set later. See
page
38.)
Setup: Horizontal
Scale: 200 ps
Position: 0 s
Reference: Center
Roll Mode: not selected
Delayed: not selected
Measure: Markers
Mode: Manual placement
All else: (n/a)
34
Setup: Channel 2
On
Scale: 50 mV/div
Offset: 500 mV
Skew: 0.0 seconds
Setup: Trigger
Mode: Edge
Source: Channel 1
Level: 500 mV
Edge: Rising Edge
Sweep: Auto
Setup: Channel 3
Off
Setup: Acquisition
Sampling Mode: Real Time
Memory Depth: Automatic
Sampling Rate: Automatic
Sine(x)/x Interpolation:
checked
Averaging: Enabled,
# of Averages: 16
16962A Logic Analyzer Service Guide
Setup: Channel 4
Off
Setup: Display
Waveforms
Connect dots: checked
Display Mode: Normal
Persistence: Minimum
Grid: On, Quantity: 1,
Intensity: as preferred