16962A Logic Analyzer Service Guide
channel. In the following example, the sampling position of
one channel (My Bus 1 [0]) must be realigned with the
sampling position of the other channels.
Using the mouse, drag the sample position (blue line) of a
1
stray channel (channel "My Bus 1 [0]" in the above
example) so that it is in the same eye as the other
channels. The Suggested Position from Eye Finder (green
triangle) will also move to the new eye.
2
Repeat the above step for all remaining stray channels.
3
Select the Run button in the Thresholds and Sample
Positions dialog. Eye Finder will recalculate the new
sample positions based on the sample position changes.
Testing Logic Analyzer Performance
3
49