Review Historical Buffer Data; Set The Data Filter - Beckman Coulter MET ONE 3400 Instructions For Use Manual

Air particle counter
Table of Contents

Advertisement

1
On the Counter Navigation screen, push
2
On the Print Center screen, select the Sample Print Mode field. Select an option for automatic
printing.
Option
None
Alarms
Cycles
All
3
Set the Cycle Print Order. Options: 10–1 (oldest printed first) or 1–10 (newest printed first).
4
To show the average of the values in the printout when the samples are taken and the data is
printed, select
NOTE
The fourth value in the printout is the average of the three previous values.
5
To round values to the nearest whole number, select
6
To repeat the header data on all samples, select
headers print only with the first sample of a series when the cycles are set to > 1.
7
Push
ENTER

Review Historical Buffer Data

Records stored in the buffer are known as Historical data. These records can be accessed
individually or sorted by location, date or time. The buffer can also be configured to collect partial
data for aborted samples.

Set the Data Filter

The data filter can be set up to sort by location, date and time, either as individual parameters or in
combination. To set the data filter, go to the Data Filter Setup screen.
DOC026.53.80202, Edition 7
Description
No data will print automatically
Prints results at the end of the sample when a count alarm is exceeded
Prints the results of multiples of the programmed count cycle. The range is 1–99.
Prints results after each count cycle is finished
.
Print Averages
to confirm.
.
PRINTER
Round Averages
Repeat Sample Headers
Operation
Review Historical Buffer Data
.
. When not selected, the
7-9
7

Advertisement

Table of Contents
loading

Table of Contents