Operation
How to Use the Test and Report Wizard

Start Sample Measurement with the Wizard

1
Push
BEGIN SAMPLING
2
Obey the prompt and go to the first sample location. Push
3
Put the isokinetic probe in position for the test. Push
4
Continue to obey the prompts and move to each location for the test. When all tests are
complete, the counter will return to the wizard to configure the report.
NOTE
With the ISO 14644-1:2015 standard, you will be prompted if you want to use sequential sampling.
The sequential sampling technique measures the rate of counting and predicts the likelihood of passing
or failing to meet the requirements of the selected ISO class. If the air being sampled is significantly more
or significantly less contaminated than the specified class concentration limit for the considered particle
size, use of the sequential sampling procedure can reduce sample volumes and sampling times, often
dramatically. When sequential sampling is selected, the wizard will automatically make this
determination and stop sampling early.

Use Existing Data

The Test and Report Wizard can include existing data for the selected standard and location.
1
Select the Use Existing Data checkbox.
2
Enter the date and time range for the existing data.
3
Push
BEGIN SAMPLING

Report Test Results

• Push
• Push
7-6
.
.
to send the selected data to a USB drive. Refer to
to print the selected data. Refer to
.
OK
.
RUN
Export
Data.
How to Use the Print
Center.
DOC026.53.80202, Edition 7