Set Or Clear The Batch Id; How To Use The Test And Report Wizard; About Standard Sampling Protocols - Beckman Coulter MET ONE 3400 Instructions For Use Manual

Air particle counter
Table of Contents

Advertisement

Operation

How to Use the Test and Report Wizard

4
Delete the Batch ID text.
5
Push
ENTER

Set or Clear the Batch ID

At the start of a sample test push
• Continue—Keep the Batch ID the same.
• Clear ID—No Batch ID is recorded with the sample cycle.
• Edit ID—Change the Batch ID for the current sample cycle.
The Batch ID can also be cleared when the sample cycle is complete.
How to Use the Test and Report Wizard

About Standard Sampling Protocols

Use the Test and Report Wizard for groups, areas and locations that require sample
measurements and reports based on specific standards and regulatory guidelines. The 3400-series
particle counters include sample strategies based on common international protocols such as EU
GMP (2015 or prior revision), ISO 14644-1:2015 or 1999, Federal Standard 209E, and British Standard
5295.
The wizard steps through seven data entry points:
• Selection of the type of standard or regulatory guideline: EU GMP (2015 or prior revision), ISO
14644-1:2015 or 1999, Federal Standard 209E, British Standard 5295, or Averages (user-defined
test protocols)
• Selection of the targeted class for the qualification of the room
• Occupancy state: As built, at rest, in operation
• Particle size or sizes to sample
• Unidirectional or non-directional air flow (for FS209E)
• Area of the room in square meters (m
• Samples per location. A minimum value is set based on room size and standard. This value can
be increased as needed.
7-4
. The default Batch ID value will show, but will not be reported.
BATCH ID
. The user can select:
2
)
DOC026.53.80202, Edition 7