5
Navigate to the needed location and select it, then push
NOTE
Push the UP or DOWN arrow to change the order of locations in the Group Settings screen. Refer
to
Change the Order of Locations
Sizes Function
IMPORTANT
maximum of ± 15% size error and a ± 30% counter error can occur when samples are taken at an
interpolated size point.
The Sizes function is a separate option that must be enabled by the manufacturer. The icon
will not show on the Counter Navigation screen if this option has not been purchased. Customized
channel sizes can be selected at time of order.
With the Sizes function enabled, the particle counter can be configured for the particle sizes that
*
follow
:
• 0.3 to 10-μm range: 0.3, 0.4, 0.5, 0.7, 0.8, 1.0, 2.0, 3.0, 4.0, 5.0, 7.0 and 10.0
• 0.5 to 25-μm range: 0.5, 0.6, 0.7, 0.8, 1.0, 2.0, 3.0, 5.0, 7.0, 10.0, 15.0 and 25.0
NOTE
Channel sizes 0.3 µm and 25.0 µm cannot be configured together.
Data Management
Store Partial Data
Data can be collected from sample measurements that have been stopped by a user or by a flow
error.
1
On the Counter Navigation screen, push
2
Select the
* Assumes 8-channel option is enabled
DOC026.53.80202, Edition 7
for more information.
If a selected channel has not been calibrated, particle counts will be based on interpolation. A
tab.
Custom
.
OK
SYSTEM.
6
Configuration
Sizes Function
6-19