About Reports

ISO 14644-1, FS 209E and BS 5295 specify calculations for airborne particle counter count data. These
documents establish the definitions for level of cleanliness in cleanrooms and clean zones based on
specified concentrations of airborne particles. The printed reports provide the data to determine
the cleanliness level for which that cleanroom qualifies.

Set up the Test and Report Wizard

1
On the Counter Navigation screen, push
2
In the Test and Report Wizard screen, complete the fields listed in this step. Select the
applicable listed options.
Option
Standard
Grade/Class
Occupancy
3
Select the Sizes field.
a. Highlight a particle size.
b. Push
c. Add as many particle sizes as needed. To remove a particle size, select the size from the
Considered Sizes list and push
4
Go to the next screen in the wizard.
a. Enter the room area in m
b. Enter the number of samples per location. Push
5
Go to the next screen in the wizard.
a. Push
Locations list when the group, area or location is added.
b. Highlight a location. Use the
The order of the list will be the order of the test.
6
Go to the next screen. A confirmation of the selections will show.
DOC026.53.80202, Edition 7
Description
Shows the list of applicable standards
Shows a list of room classification
Shows a list of room states
to add the size to the Considered Sizes list.
ADD
2
. Push
,
or
ADD AREA
ADD GROUP
UP
How to Use the Test and Report Wizard
.
TEST WIZARD
.
REMOVE
to confirm.
ENTER
to confirm.
ENTER
. Individual locations will show in the
ADD LOCATION
or
arrows to change the order of locations as needed.
DOWN