Manage The Data Buffer; Set The Data Buffer To Rotate Data; Set The Data Buffer Size - Beckman Coulter MET ONE 3400 Instructions For Use Manual

Air particle counter
Table of Contents

Advertisement

Configuration
Data Management
3
Select the
NOTE
Sample measurements collected during the execution of a Test Wizard that are stopped by a user or
flow error are always stored, regardless of the setting of this flag.

Manage the Data Buffer

The data buffer allows the data to be preserved or overwritten. In addition, the size of the buffer
can be changed and data can be cleared from the buffer.

Set the Data Buffer to Rotate Data

The default buffer setting is fixed. In this setting, no new data can be added to the buffer when it is
full. When the buffer is set to rotate, new data can be added to a full buffer, while the oldest data
record is deleted.
1
On the Counter Navigation screen, push
2
Select the
3
Select
ROTATING BUFFER

Set the Data Buffer Size

A change to the buffer size causes all current buffer data to be lost and unrecoverable.
1
On the Counter Navigation screen, push
2
Select the
3
Select the Data Buffer Size field. Enter a value between 50 and 5000.
4
Push
ENTER
5
Push
YES
6-20
checkbox.
STORE PARTIAL DATA
tab.
Custom
.
tab.
Options and Accessories
to confirm.
to clear the data buffer.
.
SYSTEM
.
SYSTEM
DOC026.53.80202, Edition 7