Emergency Scenarios
Emergency Start Ups
•
Release the oxygen safety
control knob by pressing and
set to desired oxygen flow,
between 0 and 12 l/min.
The oxygen flow is channelled
through the vaporiser.
Manual ventilation is available
at any time.
•
Switch on Primus.
After approximately 35 seconds the checklist will appear.
•
Press the cancel test key (1).
You can cancel the self test up
to 10 times. After that, it is
mandatory to perform a self test.
•
Primus switches to standby
mode after completing the
minimal test (duration 10 seconds).
•
Close safety O
•
Continue as in normal
operation.
16
flow control.
2
1