Inspection

Inspections must be carried out regularly according
to the following guidelines and within the specified
intervals. All information needed for maintenance
and safety inspections are included in this docu-
Checks
Inspection and safety checks
1) Designation applies to the Federal Republic of Germany; corresponds to the "Recurring safety inspec-
tion" in the Federal Republic of Austria

Safety inspections

CAUTION
Perform the safety inspections in the specified
intervals. Otherwise, the correct functioning of the
device may be compromised.
Scope of safety inspection for the
Cockpit
The following safety inspections are no substitute
for the inspection and maintenance indicated by the
manufacturer, including preventive exchange of
parts subject to wear. For technical documentation,
contact your local Dräger-authorized service per-
sonnel.
1 Check accompanying documents and deter-
mine if the instructions for use are available.
2 Check that the equipment is complete and
ready for use according to the instructions for
use.
3 Verify that the device in combination with other
system components is in good working condi-
tion. Specifically, verify the following:
– All labels are complete and legible
– There is no visible damage
48
Interval
1)
Every 24 months
Instructions for use - Infinity
ment and in the accompanying documents of con-
nected medical equipment. Additional guidance is
given in further technical documents that can be
obtained on request.
Responsible personnel
Service personnel
4 Check that the device meets the electrical
safety requirements according to IEC 62353,
Medical electrical equipment - recurrent test
and test after repair of medical electrical equip-
ment.
5 Verify that the visual and acoustic alarm signals
function properly.
6 Verify that the following device features operate
according to the instructions for use:
– Verify the LEDs
– Perform device checks
®
Acute Care System - Infinity
®