Remote Displays

The bedside monitor can send data to a larger VGA video display
for an enhanced view of the monitoring functions. The VGA dis-
play connects directly to the Infinity Docking Station (IDS),
interface plate, or the Communication Power Supply (CPS). Use
of a Dräger-approved video display is recommended. For order-
ing information, see the Accessories appendix.
NOTE: The remote display output on the IDS/CPS is not galvani-
cally isolated. If you use a video monitor other than the one
approved by Dräger, it must comply with IEC 60601-1. Upon
installation, the installer must ensure that in normal and single fault
conditions, the entire system meets the requirements of IEC 60601-1
and IEC 60601-1-1 (Medical Electrical Systems Standards). Refer
to the video monitor's operating instructions to ensure that the
interconnection is within its intended use as specified by the manu-
facturer. Additionally, radiated and conducted emissions classifica-
tion, suitability for flammable locations and water ingress
protection must be considered based on the intended use of the sys-