Testing Readiness for Operation
Checking LEDs, displays and audible alarm
1 Press the
key.
Check
2 The following message appears in the display:
»All displays on, horn on«.
All the displays are lit for about 2 seconds:
digital displays read »88.8« and a continuous tone sounds.
2 The following message appears in the display:
»All displays off, soft alarm on«.
All displays are dark for about 2 seconds and the alarm
sounds. A function check can also be carried out while the
unit is in operation.
●
Check at least once a day.
Testing the power failure alarm
Disconnect the power plug.
●
3 The red LED
should light up and a continuous alarm
tone should sound.
Reconnect the power. The unit will continue to operate
●
with the values set before the power failure alarm.
Testing the lights
4 Press the left-hand button. The bed should now be illumi-
nated by the procedure light.
4 Press the button again. The procedure light will be
switched off.
5 Press the middle button. The bed should now be lit by the
night light.
5 Press the button again. The night light will be switched off.
Testing phototherapy (optional)
6 Press button. The yellow LED in the button will start to
flash.
8 The following message appears in the text display:
»Phototherapy: Use eye protection
Press OK button to start
7 Press OK button = phototherapy lights switched on.
8 The following message appears in the text display for
5 seconds:
»Duration of phototherapy
6 The yellow LED in the button is now continuously lit.
6 Press button.
8 The following message appears in the text display:
»Phototherapy turned OFF
Press OK button to confirm XX:XX:XX«.
7 Press OK button = phototherapy switched off.
6 The yellow LED in the button will go out.