Checking readiness for operation
Checking ventilation

Checking end expiratory pressure PEEP

Alternating between inspiration and expiration, the
pressure gauge should indicate a defined inspiratory
pressure and the end expiratory pressure of approx.
5 mbar.
The display shows a minute volume of
IPPV MV = 6
with a tolerance of ±1.0 L/min.
If the measured value of MV is only 5 L/min:
Check that the ventilation hose, breathing valve,
flow sensor, elbow connector and test lung are
undamaged and connections are tight.
If the measured value of MV = 6 L/min deviates by more
than 1 L/min:
Replace flow sensor.
Checking end expiratory pressure PEEP
1 Set »PEEP« knob to 0 mbar.
2 Pressure gauge display at end of expiration:
0 mbar +2 mbar
1 Set »PEEP« knob to 10 mbar.
2 Pressure gauge display at end of expiration:
10 mbar ±2 mbar
1 Set »PEEP« knob back to 0 mbar.