To skip the Welcome Menu:
Press VENTILATION ON/OFF
The Ventilation menu is then displayed.
By default, the starting ventilation mode is the last one used, the settings being those that were
active when the machine was last stopped.If the ventilator's memory of the settings is faulty, a
"CHECK SETTINGS" alarm is activated. If this occurs, the desired parameters should be reset and
saved; otherwise the machine will operate on default parameter values.
5.2

USB Menu Parameters

The USB menu is accessible even if the Locking Key has been enabled. The USB menu is automatically dis-
played when the USB memory device is connected to the ventilator, when ventilation is either on or off.
Only one USB memory device shall be connected at any time, otherwise an error message will be displayed.
The USB Menu is not accessible from the Setup Menu or Maintenance menu.
To access patient data via a PC, a dedicated software package, Puritan Bennett™ Respiratory Insight Software,
is available for Clinicians. Contact Covidien or your Puritan Bennett™ product representative for further infor-
mation.

5.2.1 USB Memory Device Specifications

Table 5-1. USB Memory Device Specifications

Characteristics
USB compatibility
Number of files
USB size
User's Manual
to start ventilation immediately.

Figure 5-3. Ventilation Menu Parameters

Supported Formats
USB flash memory USB 2.0 or USB 1.1, 32 bit format
Maximum 999 (sector size: 512-2,048 bytes)
128 MB to 4Gb (To guarantee accuracy of transfer time, at
least 10% of the USB memory device capacity must be free).
Operating Procedures
5-3