Operating Parameters

Ambient temperature
range
Relative humidity
Atmospheric pressure
Warm-up time

Transport and Storage

Ambient temperature
range
Relative humidity
Atmospheric pressure
Duration of storage

Duty Cycle

Under maximum-output settings and rated-load conditions (30 ohm load), the generator
is suitable for activation times of 5 seconds on, 15 seconds off, for 1 hour. With lesser
settings and loads, you can activate the generator for greater durations without
generating excessive internal temperatures.

Internal Battery

Battery for RTC
Valleylab LS10, LS Series Single Channel Vessel Sealing Generator
+10° C to +40° C (50° F to 104° F)
30% to 75% non-condensing
700 to 1060 millibars
If transported or stored at temperatures outside
the operating temperature range, allow one hour
for the generator to reach room temperature
before use.
-30° C to 65° C (-22° F to 149° F)
25% to 85% (non-condensing)
500 to 1060 millibars
If stored over one year, see service manual for
instructions or contact Covidien Service for further
information.
Battery type – 3 V lithium button cell
Battery life – 5 years
Performance Characteristics