Notes On Operation - Beckhoff EPP3504-0023 Short Manual

4-channel measuring bridge,(sg) full/half/quarter bridge, 24 bit, 10 ksps
Hide thumbs Also See for EPP3504-0023:
Table of Contents

Advertisement

Appendix
8.7

Notes on operation

EMC immunity according to EN 61000-6-2
In order to achieve immunity to electromagnetic interference in accordance with EN 61000-6-2, a digital low-
pass filter must be used with which the peak-to-peak (PtP) noise is suppressed to maximally 1/3 of the
specified basic accuracy. Such filtering is also useful from the point of view of measurement technology, as
the basic accuracy given for the respective terminal is specified "with averaging", i.e. without the influence of
white noise. Filtering is not necessary if the unfiltered PtP noise already lies below the 1/3 limit. It may also
be useful to do without filtering when fast response times are required. However, it is then possible that the
full specified basic accuracy is not achievable in all cases.
Here are two examples of this with fictitious numbers:
• Example 1:
ELM3004-0000, ±10 V measuring range:
◦ Noise without filtering E
◦ Because 60 < (1/3 ⋅ 200), no filtering is required.
• Example 2:
ELM3004-0000, ±20 V measuring range:
◦ Noise without filtering E
◦ Now 560 > (1/3 ⋅ 600), and a low-pass filter is required if the full basic accuracy is to be achieved.
244
< 60 ppm
, basic accuracy ±100 ppm
Noise,PtP
FSV
< 560 ppm
, basic accuracy ±300 ppm
Noise,PtP
FSV
Version: 1.2
.
FSV
.
FSV
EPP3504-0023