Beckhoff EPP3504-0023 Short Manual page 74

4-channel measuring bridge,(sg) full/half/quarter bridge, 24 bit, 10 ksps
Hide thumbs Also See for EPP3504-0023:
Table of Contents

Advertisement

Commissioning
Index
Name
(hex)
90n2:01 Effective Sample
Rate
90n2:02 Channel
Temperature
90n2:03 Min. Channel
Temperature
90n2:04 Max. Channel
Temperature
90n2:05 Overload Time
90n2:06 Saturation Time Absolute time during saturation
90n2:07 Overtemperature
Time (Channel)
90n2:10 Load Cycle
Counter
90n2:11 Vendor
Calibration
Counter
90n2:12 User Calibration
Counter
74
Meaning
Effective Sample Rate
Temperature of the channel
Minimum temperature of the channel
Maximum temperature of the channel
Absolute time during overload
"Overload" means that the channel is
electrically overloaded. This is a non-
recommendable condition that may
eventually lead to atypical aging or
even damage. This condition should be
avoided.
Its accumulated duration is displayed
here informatively.
"Saturation" means that the measuring
range of the ADC of the channel is fully
utilized, the ADC thus outputs its
maximum value and the measured
value can no longer be used.
"Saturation" is therefore a pre-
deregistration, with further signal
increase it comes to "overload".
The saturation state is not
fundamentally harmful, but it indicates
an insufficient dimensioning of the
measurement channel.
Its accumulated response time is
displayed here informatively.
Time of exceeded temperature of the
channel
Load Cycle Counter
Counter of the vendor calibration
(related to the selected interface)
The counter counts +1 when data has
changed and the memory code word is
written. Depending on the adjustment
method, the counter may therefore
count several times.
Counter of the user calibration
(related to the selected interface)
The counter counts +1 when data has
changed and the memory code word is
written. Depending on the adjustment
method, the counter may therefore
count several times.
Version: 1.2
Data type
Flags
UINT32
RO
REAL32
RO
REAL32
RO
REAL32
RO
UINT32
RO
UINT32
RO
UINT32
RO
UINT32
RO
UINT16
RO
UINT16
RO
Default
0x00000000
(0
)
dec
0x00000000
(0
)
dec
0x00000000
(0
)
dec
0x00000000
(0
)
dec
0x00000000
(0
)
dec
0x00000000
(0
)
dec
0x00000000
(0
)
dec
0x00000000
(0
)
dec
0x0000
(0
)
dec
0x0000
(0
)
dec
EPP3504-0023