Maintenance
Maintaining Your Instrument
24
Nicolet iS5N Spectrometer User Guide
Before you begin, make sure the spectrometer has been powered on for at least 15 minutes
(one to six hours for best results).
To run performance verification
1. Make sure the iD1 transmission accessory is installed in the spectrometer sample
compartment.
2. Insert the C screen from the iS5N accessory kit into the card slot furthest to the right
inside the iD1 sample compartment.
(There should not be a sample in the iD1 sample compartment.)
3. Click the System Status shield in the OMNIC main window.
The System Status Overview box is displayed.
4. Choose Performance Verification and then choose Run.