Aligning Your Spectrometer

Verifying the Laser Frequency

Thermo Scientific
Related Topics
Installing a Purge Kit
Checking and Changing the Purge Gas Filter
The spectrometer can be aligned automatically using OMNIC software. Alignment optimizes
the energy throughput to the detector. Alignment can change slightly over time, particularly
in locations that have large temperature swings or excessive vibration.
To align the spectrometer
1. From the OMNIC software main screen, choose Collect > Experiment Setup.
2. Select the Diagnostic tab and click Align.
For more information, see OMNIC Help Topics in the Help menu.
Note The spectrometer is aligned automatically each time you run a Performance
Verification (PV) test. We recommend running a PV test at least once a month.
Related Topics
Verifying the Laser Frequency
Maintaining Your Instrument
Troubleshooting
Use the Laser Verification button on the Diagnostic tab in the Experiment Setup box in
OMNIC software to verify the laser frequency. For optimum system performance, verify the
laser frequency according to the following schedule:
Spectrometer environment
Stable
Unstable (e.g., temperature or
humidity constantly changing)
Note The laser is verified automatically each time you run a Performance Verification
(PV) test. We recommend running a PV test at least once a month.
Spectrometer state
Always powered on
Powered on and off
frequently
Nicolet iS5N Spectrometer User Guide
2
Operation
Aligning Your Spectrometer
Recommended Schedule
every 3 months
every day or once a week
17