Regenerating the Desiccant

Thermo Scientific
Click the Installation icon. A diagnostic information screen appears.
Enter today's date for the desiccant replacement and click Save.
15. Reconnect the purge line to the instrument, if applicable, and turn on the purge.
16. Make sure the system has been on for at least 15 minutes (one to six hours for best
results), and then run the
17. Install any accessories you removed earlier and reconnect the purge, if applicable.
Related Topics
Checking the Humidity Indicators
Regenerating the Desiccant
Installing a Purge Kit
When the desiccant canisters must be replaced, you can
canisters, or you can dry and reuse the saturated canisters according to the following
instructions.
Time needed: 3 hours
Tools needed:
• Used desiccant canisters
• Vented oven
• Insulated cloth or hot pad
Performance Verification
3
Maintaining Your Instrument
(PV) test.
contact us
to order new desiccant
Nicolet iS5N Spectrometer User Guide
Maintenance