5. Remove and discard the large O-ring.
Note Be careful not to drop the sealing O-ring into the desiccant compartment.
6. Remove the new O-ring from the kit and install it on the instrument base plate (see the
previous image).
Press (rather than roll) the new O-ring into place to be sure it is properly seated in the
groove.
7. Use the 0.05-inch Hex key to loosen (approximately two turns) the eight screws that hold
the two saturated desiccant canisters to the desiccant cartridge. Remove the canisters.
Note The desiccant canisters can be regenerated. See
details.
Humidity indicator
Desiccant canisters
3
Maintenance
Maintaining Your Instrument
Regenerating the Desiccant
Desiccant cartridge
Channel
Remove these screws
(4 on each side)
Nicolet iS5N Spectrometer User Guide
for
31