Assembly and preparation

Preparing the ventilator

Preparing the expiratory valve

WARNING
Risk of patient injury
Expiratory valves that are damp or have not
been reprocessed can impair the operation of
the device and endanger the patient.
Only use properly reprocessed expiratory
valves which have been sufficiently dried.
CAUTION
High airway pressures and auto-triggering
If the water trap container on the expiratory valve
is missing, there is a danger of excessively high
airway pressures and auto-triggering due to leak-
age overcompensation.
Always attach the water trap container.
Assembling the expiratory valve
1 Fit the diaphragm (A) onto the edge of the
expiratory valve housing.
2 Make sure that the diaphragm is fitted properly.
3 If the flow sensor sleeve (B) has been removed,
fit the flow sensor sleeve.
52
B
A
4 Attach the water trap container (C).
Opening the flap
Open the flap (D) before inserting the expiratory
valve.
D
 Lift the flap (D) by the lower edge and pivot it
upwards.
Instructions for use Savina 300 SW 5.n