Replacing The Dust Filter Set - Dräger Savina 300 Instructions For Use Manual

Hide thumbs Also See for Savina 300:
Table of Contents

Advertisement

Replacing the dust filter set

CAUTION
Risk of malfunction
Soiled dust filters may impair the proper
functioning of the device.
Replace the dust filter set at regular intervals.
Visually check the dust filter set for soiling after
4 weeks and clean or replace as necessary.
Replace after 1 year at the latest.
1 Remove the filter cover, see page 67.
A
2 Remove the soiled dust filter set (A) from the
filter cover (B) and dispose of with domestic
waste.
3 Insert the new dust filter set (A).
4 Fit the filter cover, see page 67.
Instructions for use Savina 300 SW 5.n
A
B
A
Service
191