Reprocessing procedures

Manual cleaning followed by disinfection by
immersion
The following components are suitable for manual
cleaning followed by disinfection by immersion:
– Expiratory valve
The following disinfectants were compatible with
the material at the time of testing:
Cleaning agent:
neodisher LM2
Disinfectants
Korsolex Extra
Steam sterilization
The following components can be steam sterilized:
– Expiratory valve
NOTE
Sterilize the components of the expiratory valve
(valve housing, flow sensor sleeve, diaphragm,
water trap container) only after disassembly.
Prerequisites:
– The components have been cleaned,
disinfected, and are dry.
– Use a steam sterilizer that meets the
requirements of the standard ISO 17665.
Dräger recommends steam sterilization with
fractionated vacuum.
Procedure:
1 Sterilize the components (maximum 134 °C
(273.2 °F), 5 min).
2 Check the components for visible damage and
replace if necessary.
Instructions for use Savina 300 SW 5.n
Manufacturer
Dr. Weigert
BODE Chemie
Reprocessing
183