Service

Safety information

WARNING
Risk if service is not performed regularly
Wear and material fatigue of the components
may lead to device failure and malfunctions.
Perform service at the specified intervals.
WARNING
Risk if service is not performed properly
Personal injury and property damage may
occur if service is not performed properly.
Service must be performed by those user
groups that are assigned to the particular
measure.
WARNING
Risk when the housing is being opened
Under the housing, there are live electrical
components, which may cause an electric
shock.
The housing may only be opened by those
user groups that are assigned to that
particular measure.
CAUTION
Risk if the battery is not replaced properly
If the battery is not replaced properly, short
circuits and high temperatures leading to
explosion or fire may occur.
The battery must be replaced by the assigned
user groups.
186
CAUTION
Risk due to inappropriately reprocessed products
The product may be contaminated with infectious
agents.
Before service is performed and before the
product is sent back for repair, reprocess the
product in accordance with the chapter
"Reprocessing".
CAUTION
Risk of medical device failure
If the safety-relevant tests are not performed on a
regular basis, the proper functioning of the
medical device can be compromised.
The electrical safety test and the functional tests
described in the technical documentation IPM
must be carried out within the specified intervals.
NOTE
If Savina 300 is used under the following
conditions, the specified intervals, e.g., for
checking the batteries, must be reduced:
– Environments of use with extreme conditions
– Frequent intrahospital patient transports
Reduce the intervals after consulting
DrägerService.
Instructions for use Savina 300 SW 5.n