Technical data
Performance characteristics (cont.)
Device compliance
(with bacterial filter, 2.7 to 2.8 m (8.9 to 9.2 ft)
breathing circuit for pediatric patients, Fisher &
Paykel breathing gas humidifier MR 850 with emp-
ty F&P humidifier chamber MR 340, breathing
hoses heated or unheated and water traps)
Maximum values for the combination of accesso-
ries in accordance with ISO 80601-2-12 (including
inspiratory bacterial filter)
Breathing circuit for adults
Compliance
Inspiratory resistance
Expiratory resistance
Breathing circuit for pediatric patients
Compliance
Inspiratory resistance
202
1 mL/mbar
1 mL/hPa
1 mL/cmH
O
2
3,2 mL/mbar
3,2 mL/hPa
3,2 mL/cmH
O
2
16 mbar/L/s at 60 L/min
16 hPa/L/s at 60 L/min
16 cmH
O/L/s at 60 L/min
2
12 mbar/L/s at 30 L/min
12 hPa/L/s at 30 L/min
12 cmH
O/L/s at 30 L/min
2
10 mbar/L/s at 60 L/min
10 hPa/L/s at 60 L/min
10 cmH
O/L/s at 60 L/min
2
10 mbar/L/s at 30 L/min
10 hPa/L/s at 30 L/min
10 cmH
O/L/s at 30 L/min
2
1 mL/mbar
1 mL/hPa
1 mL/cmH
O
2
50 mbar/L/s at 30 L/min
50 hPa/L/s at 30 L/min
50 cmH
O/L/s at 30 L/min
2
40 mbar/L/s at 15 L/min
40 hPa/L/s at 15 L/min
40 cmH
O/L/s at 15 L/min
2
10 mbar/L/s at 2.5 L/min
10 hPa/L/s at 2.5 L/min
10 cmH
O/L/s at 2.5 L/min
2
Instructions for use Savina 300 SW 5.n