Technical data
Set values (cont.)
Pressure
Inspiratory pressure P
Positive end-expiratory pressure PEEP or in-
termittent PEEP
Pressure support
Inspiratory pressure limit P
Sigh pressure
intPEEP
Accuracy
Trigger sensitivity
Can be switched off in VC-CMV / VC-AC ventila-
tion mode
Trigger switched off = VC-CMV
Trigger switched on = VC-AC
Accuracy
Termination criterion (peak inspiratory flow)
Insp. term.
Accuracy
Flow acceleration FlowAcc
Accuracy
Airway Pressure Release Ventilation
T
high
T
low
Accuracy
Resolution
P
high
P
low
Accuracy
Automatic tube compensation
Tube type
Inner diameter of the tube
200
insp
P
above PEEP
supp
max
1 to 99 mbar (or hPa or cmH
[1 mbar = 100 Pa]
0 to 50 mbar (or hPa or cmH
0 to 50 mbar (or hPa or cmH
(relative to PEEP)
1 to 99 mbar (or hPa or cmH
[1 mbar = 100 Pa]
0 to 20 mbar (or hPa or cmH
±2 mbar (or hPa or cmH
1 to 15 L/min
±8 % of the set value or ±0.5 L/min, whichever is
greater
5 to 75 % PIF
Dependent on the accuracy of flow measurement
5 to 200 mbar/s (or hPa/s or cmH
±20 % of the set value
APRV
0.2 to 22.0 s
0.1 to 22.0 s
0.1 s
0.05 s in the range up to 1 s, 0.1 s in the range
above 1 s
1 to 95 mbar (or hPa or cmH
0 to 50 mbar (or hPa or cmH
±2 mbar (or hPa or cmH
ATC
Endotracheal tube ET or tracheostomy tube Trach.
3.5 to 12.0 mm in increments of 0.5 mm
Instructions for use Savina 300 SW 5.n
O)
2
O)
2
O)
2
O)
2
O)
2
O)
2
O/s)
2
O)
2
O)
2
O)
2