Technical data
Displayed measured values (cont.)
Mean airway pressure
Range
Resolution
Accuracy
Intrinsic PEEP
Externally applied PEEP
Range
Resolution
Accuracy
Inspiratory O
measurement
2
(maintenance-free electrochemical sensor, ambi-
ent pressure-compensated)
Inspiratory O
concentration
2
Range
Resolution
Accuracy in HPO mode and LPO mode
Drift of measurement accuracy
Warm-up time
Maximum response time from 21 Vol% to
60 Vol%
Flow measurement
Constant flow (O
therapy)
2
Inspiratory peak flow
Range
Resolution
Accuracy
Minute volume measurement
Minute volume
Spontaneous minute volume
Range
Resolution
Accuracy
T
10...90
204
P
mean
0 to 99 mbar (or hPa or cmH
1 mbar (or hPa or cmH
±2 mbar (or hPa or cmH
PEEPi
incl. PEEP
0 to 100 mbar (or hPa or cmH
0.1 mbar (or hPa or cmH
±2 mbar (or hPa or cmH
FiO
2
18 to 100 Vol%
1 Vol% O
2
±3 Vol% O
2
±1 Vol%/day
10 to 20 min
T
<20 s
0...90
Flow
Flow
ipeak
0 to 196 L/min BTPS
1 L/min
±10 % of the measured value or ±1 L/min,
whichever is greater
MV
MV
spon
0 to 99 L/min BTPS
0.01 L/min in the range up to 10 L/min
0.1 L/min in the range above 10 L/min
±12 % of the measured value or ±0.6 L/min,
whichever is greater
Approx. 35 s
Instructions for use Savina 300 SW 5.n
O)
2
O)
2
O)
2
O)
2
O)
2
O)
2