Interrupting The Therapy - Standby Mode; Activating Standby Mode - Dräger Savina 300 Instructions For Use Manual

Hide thumbs Also See for Savina 300:
Table of Contents

Advertisement

Operation
Interrupting the therapy – standby mode
If the standby mode is activated, ventilation is
interrupted. Switch to standby mode for the
following actions:
– Keeping Savina 300 ready for operation while
the patient is absent
– Changing the application mode
– Changing the therapy type
– Performing a device check
WARNING
Risk of patient injury
Ventilation does not take place in standby
mode. Patients connected to the device are
endangered.
Only set the device to standby mode when no
patient is connected to the device.

Activating standby mode

1 Press the
Start/Standby key.
Savina 300 opens the Start/Standby (A) page.
A
B
2 Touch the Standby button (B) and confirm with
the rotary knob.
The alarm message Standby mode activated is
displayed in the header bar.
3 Press the
Alarm reset key.
104
Savina 300 is in standby mode. Standby is
displayed in the header bar of the screen instead of
the ventilation mode. Information on the last
ventilation settings continues to be displayed, e.g.,
AutoFlow.
Additional information
For information on continuing the therapy, see
"Starting the therapy" on page 82.
Instructions for use Savina 300 SW 5.n