Performing the coaxial breathing circuit

check

For coaxial breathing circuits, the leakage of the
inner hose in measured as well.
Preparing the coaxial breathing circuit check
Prerequisites:
– Savina 300 is prepared and in standby mode.
– The Breathing circ. check page (A) is opened.
– The breathing circuit Coaxial (C) is selected,
see page 72.
A
B
C
E
F
H
I
The overall result of the last check with date and
time is displayed in the field (G).
The procedure is described in the instruction field
(H). The graphics (D) show the respective test
setup.
Instructions for use Savina 300 SW 5.n
K Inspiratory port
L Expiratory port
M Airway connector
D
N Catheter mount
O HME/filter
G
P Patient port of the coaxial breathing circuit
Q Red test adapter
J
1 Connect the inspiratory connector of the coaxial
2 Connect the expiratory connector of the coaxial
3 Connect the HME/filter, catheter mount, and
Starting the coaxial breathing circuit check
1 Touch the Start button (E).
2 Confirm with the rotary knob.
3 Remove the HME/filter, catheter mount, and
4 Plug the red test adapter (Q) onto the inner
5 Seal the red test adapter, e.g., with a sterile
6 Confirm with OK (I).
The leakage test for the inner hose is performed.
Q
breathing circuit to the inspiratory port (K).
breathing circuit to the expiratory port (L).
airway connector to the patient port of the
coaxial breathing circuit (P).
airway connector.
hose of the coaxial breathing circuit (P).
glove.
Getting started
K
L
M
N
O
P
77