AV-E ANESTHESIA VENTILATOR

The anesthesia ventilator (Figure 13) is
volume-preset and time-cycled. It has a
solid-state timer and independent controls.
The pneumatic power to the ventilator may
be supplied through the pipeline supply or,
if the pipeline supply either fails or is
disconnected,
through
pressure of the supply gas must be
FREQUENCY
CONTROL
TIDAL VOLUME
ADJUSTMENT
KNOB
VENTILATOR RELIEF
VALVE (W/19MM
SCAVENGER
HOSE TERMINAL)
22MM BREATHING
HOSE TERMINAL
Figure 13
AV-E Ventilator
20
cylinders.
The
I.E.
GAUGE
RATIO
CONTROL
RETURN TO CD-ROM TABLE OF CONTENTS
between 40 and 60 psi. The ventilator will
not function if this pressure drops below 32
psi.
The
monitoring
pressure and expiratory flow waveform
displays can be used as an aid in adjusting
the ventilator. For further details, see the
MONITORING SYSTEM section of this
manual.
FLOW
FLOW
CONTROL
system's
breathing
VENTILATOR
POWER
SWITCH
SETTING POINTER
TIDAL VOLUME
OP30014